3. 2. 1測試姿勢Position:
水平放于治具上Hoizontal
3. 2. 2用500 cc容器承受空氣壓力300 mmHg
Ai pessue 300mmHg with 500CC tank
3. 2. 3設定電壓規格Rated Voltage:
DC 6.0V
使用電壓范圍Opeating Voltage Range:
DC4.2-7.2V
3. 2. 4設定電流Rated Cuent:
低于450mA Less than 450mA
使用設定電壓供應幫浦充氣從0到300mmHg的最高電流.
Duing pessuization fom 0-300mmHg at ate voltage.
3. 2. 5充氣時間Inflation time:小于8.0秒Less than 8.0 sec
在500cc容器中充氣壓力從0到300mmHg所需時間.
To pessuize in a 500cc tank fom 0-300mmHg at ated voltage
3. 2. 6漏氣Leakage:小于10mmHg/min. Less than 10mmHg/min
在500cc容器中充氣壓力從0-300±30mmHg之后停止充氣,安定30秒后
測其一分鐘減少的壓力值.
Pessuizing a 500cc tank fom 0-300±30mmHg. Pessue dop shall be less than
10mmHg / min in 30 sec afte the stop of the pump.
3. 2. 7.噪音Noisy:少于55dB Less than 55dB.
將邦浦放于密閉箱內,距噪音計30cm之處,置于一5CM厚的海綿上,充氣從0到
300mmHg所測得dB之值. (圖1)
Put the pump into pessuized box and distance to noise mete 30cm, and while pessuing fom 0 to 200 mmHg to test the noise value. (As fig 1)
3. 2. 8最大氣壓Maximum Pessue:大于500 mmHg moe than 500mmHg
以設定電壓在500cc容器中充氣壓力從0到停止升壓之最高壓力值
Pessuize at a 500cc tank. The maximum pessuization value is while the pump stops the inflation.
3. 2. 9壽命測試Life Test:
設定充氣8秒,停止8秒不間斷測試30,000次以后經測試無異常,漏氣量需低于
18mmHg/分鐘,噪音低于72dB
Afte pefoming opeation 30,000 times in the state of set up 8secinflating, then stop 8sec. Thee ae no abnomalities in a deed above test. Leakage shall be less than 18mmHg/min, and noise less than 72dB.本公司创立于1976年,历经30年的稳定.踏实成长,产品经过不断创新.改进.开发出数十项专利产品:有医疗.环保.运动.休闲.血压计气泵.保健.水泵。在美国.日本.德国.中国.台湾取得专利证书及商标权,行销世界各地,对于品质的提升,客户的满意.要求,为达到最高品质目标,已正式通过图际ISO9001:2000及医疗级13485质量管理体系认证,未来计划更多的国际认证及产品的再创新而。开发能力:欢迎新产品开发OEM,ODM。模具制作:模具设计.模具制造.产品设计.产品制图。注塑成型:ABS.PP.PS.PVC.尼龙加纤。产品:打气泵,真空(吸力)泵,水泵。打气泵真空泵:医疗用,环保用,臭氧用,机械用,钓具用,气床用,玩具用,吹气用,血压计用,保健用,喷气用,抽痰机